Battery Saver

The Calibrator automatically turns off after 30 minutes of
inactivity. To reduce this time or disable this feature:
1.
With the Calibrator OFF, press
P.S. xx is displayed, where xx is the turn-off time in
minutes. OFF means the power saver is disabled.
2.
Turn
to decrease or
To disable, turn
until the display shows OFF.
3.
The Calibrator resumes normal operation after
2 seconds.

Safety Information

In this manual a Warning identifies conditions and actions
that pose hazards to the user. A Caution identifies
conditions and actions that may damage the Calibrator or
equipment under test. International symbols used in this
manual are identified later in the Symbols section.
Read the entire Users Manual and the 707Ex mA
Calibrator CCD (Concept Control Drawing) before using
the Calibrator.
2
.
to increase the turn-off time.

WWarnings and Cautions

To avoid electric shock, injury, or damage to the
Calibrator:
Use the Calibrator only as described in this
Users Manual and the Fluke 707Ex mA
Calibrator CCD (Concept Control Drawing) or
the protection provided by the Calibrator may
be impaired.
Inspect the Calibrator before use. Do not use it
if it appears damaged.
Check the test leads for continuity, damaged
insulation, or exposed metal. Replace damaged
test leads.
Never apply more than 28 V between the input
terminals, or between any terminal and earth
ground.
Applying more than 28 volts to the input
terminals invalidates the Calibrator's Ex
Approval and may result in permanent damage
to the unit so it can no longer be used.
Use the proper terminals, mode, and range for
your measuring or sourcing application.
To prevent damage to the unit under test, put
the Calibrator in the correct mode before
connecting the test leads.
When making connections, connect the COM
test probe before the live test probe. When
disconnecting, disconnect the live probe before
the COM probe.