Should Agilent and the user have a separate written agreement with warranty terms covering the material in this WA RN ING document that conflict with these...
Description The 85130C adapters are used when type-N devices must be attached to a 3.5 mm test set in a network analyzer system. The test set end of the adapters has a NMD-3.5 mm (f) connector while the Device Under Test (DUT) end has a precision slotless type-N connector (PSC-N).
Specifications Agilent Technologies guarantees that your adapters will equal or exceed the following specifications in the to +26 C (+68 to +79 F) temperature range. Frequency Return Loss 34 dB DC – 8 GHz 28 dB 8 –...
Analyzers Configuration Guide, time domain option 010 part number 5990-7745EN) Type-N loads (included in the 00909-60011 (male) 85054B calibration kit) 00909-60012 (female) Type-N 50 ohm airline, 12.5 cm 85055-60001 (included in the 85055A verification kit) 85130C Operating and Service Manual...
Page 6
8. Set a stimulus stop time for the sweep to 1.9 nano-seconds. 9. Select the gating function and gate-out everything but the adapter. See Figure 10. Select the analyzer’s frequency domain mode. 11. Use the markers to read the return loss value. 85130C Operating and Service Manual...
NMD-3.5 mm (f) to PSC-N (m) 28 dB 8 GHz to 18 GHz NMD-3.5 mm (f) to PSC-N (f) Replaceable Parts There are no replaceable components for the adapters. A worn or damaged adapter must be replaced in whole. 85130C Operating and Service Manual...
8710-1764 Document: Connector Care for RF and Microwave Coaxial Connectors 08510-90064 Contacting Agilent Assistance with test and measurement needs and information on finding a local Agilent office are available on the Web at: www.agilent.com/find/assist NOTE In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number.