HP ESA-L1500A User's And Calibration Manual page 246

Spectrum analyzer
Table of Contents

Advertisement

Figure 9-13
Frequency Response Test Setup,
Use only 75
CAUTION
connectors, or damage to the connectors will occur.
3. Connect the equipment as shown in
4. Press
5. Adjust the synthesized signal generator amplitude for 0 dBm
6. Record the power meter reading in column 4 of
7. On the synthesized signal generator, press
8. Repeat
9. On the synthesized signal generator, press
Chapter 9
cables, connectors, or adapters on instruments with 75
mechanical adapter on analyzers with 75
,
Blue Key
Special 00
the synthesized signal generator controls as follows:
,
FREQUENCY
1 MHz
,
INCREMENT SET
50 MHz
,
AMPLITUDE
5 dBm
reading on the measuring receiver.
into account the calibration factors of both the 100 kHz to 1500 MHz
power sensor and the 75
step 5
and
step
factor into the respective power meter.
(up-arrow key), to step through the remaining frequencies listed in
Table
9-15. (No measurement is made at 100 kHz).
100 kHz
Figure
9-12. Be sure to use a
on the synthesized signal generator. Set
power sensor.
FREQUENCY
6, entering each power sensor's calibration
FREQUENCY
Performance Verification Tests
13. Frequency Response
inputs.
Table
9-15, taking
,
.
50 MHz
, and
245