18. Calibrator
Amplitude
Accuracy Test
Related Adjustment
Specification
Description
Equipment
Procedure
2-62 Performance Tests
20 MHz Reference Adjustments
-10 dBm f0.3 dB
The output level of the calibrator signal is measured with a power
meter.
Figure 2-32. Calibrator Amplitude Accuracy Test Setup
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 436A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8482A
Adapter, Type N (f) to BNC (m) . . . . . . . . . . . . . . . . . . . . . . . . HP 1250-0077
1. Connect equipment as shown in Figure 2-32.
2. Measure output level of the CAL OUTPUT signal. The value should
be -10.0 dBm ho.3 dB.