Execution Errors - Fluke 8508A Service Manual

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

8508A
Service Manual
9006 - Serial Loop Mismatch
xxxxxxxx is the actual data received in binary format.
See Table 4-7 for descriptions of the serial loop mismatch errors.
9007 Unknown Engine Instruction Addr id
id indicates the code that the engine was attempting to interpret as an instruction.

Execution Errors

An Execution Error is generated if a received command cannot be executed because it is
incompatible with the current device state, or because it attempts to command parameters
which are out-of-limits.
In remote operation, the EXE bit (4) is set true in the Standard-defined Event Status Byte,
and the error code number is appended to the Execution Error queue.
The error is associated with status reporting, and the queue entries can be read
destructively as LIFO by the Common query command EXQ?. There is no queue when
execution errors occur during manual operation; instead, the description of the error is
sent directly on the Menu display.
See Table 4-8 for descriptions of the execution errors.
4-18
Table 4-7. Serial Loop Mismatch Errors
id
00
U901_C pattern mismatch – should be 00000001 (00000000 if rear fitted)
01
U901_B pattern mismatch – should be 00010000
02
U901_A pattern mismatch – should be 10100100
03
U902_C pattern mismatch – should be 00000000
04
U902_B pattern mismatch – should be 00000000
05
U902_A pattern mismatch – should be 00000000
06
U903_C pattern mismatch – should be 00010101
07
U903_B pattern mismatch – should be 01111111
08
U903_A pattern mismatch – should be 11111011
xxxxxxxx id
Fault

Advertisement