Good Laboratory Practice (Glp) - Hanna Instruments HI 98180 Instruction Manual

Calibration check waterproof ph/mv/ise/temperature meters
Table of Contents

Advertisement

GOOD LABORATORY PRACTICE (GLP)

GOOD LABORATORY PRACTICE (GLP)
GOOD LABORATORY PRACTICE (GLP)
GOOD LABORATORY PRACTICE (GLP)
GOOD LABORATORY PRACTICE (GLP)
GLP is a set of functions that allows storage and retrieval of data regarding the
maintenance and status of the electrode.
All data regarding pH, Rel mV or ISE calibration is stored for the user to review
when necessary.
EXPIRED CALIBRATION
The instrument is provided with a real time clock (RTC), in order to monitor
the time elapsed since the last pH calibration.
The real time clock is reset every time the instrument is calibrated and the
"expired calibration" status is triggered when the instrument detects a
calibration time out. The "CAL" "DUE" tags will start blinking to warn the
user that the instrument should be recalibrated.
The calibration time out can be set (see SETUP for details, page 33) from 1 to
7 days or can be disabled.
For example, if a 4 days time out has been selected, the instrument will issue
the alarm exactly 4 days after the last calibration.
However, if at any moment the expiration value is changed (e.g. to 5 days),
then the alarm will be immediately recalculated and appear 5 days after the
last calibration.
Notes: • When the instrument is not calibrated or calibration is cleared
(default values loaded) there is no "expired calibration", and
the display always shows the "CAL" "DUE" tags blinking.
• When an abnormal condition in the RTC is detected, the instrument
forces the "expired calibration" status.
LAST pH CALIBRATION DATA
The last pH calibration data is stored automatically after a successful calibration.
To view the pH calibration data, press GLP when the instrument is in the pH
measurement mode.
The instrument will display a lot of data including calibration buffer, offset,
slope, electrode condition.
Note: Buffers displayed in video inverse mode are from previous calibrations.
The custom buffers are marked with an "
"No user calibration" message is displayed if all calibration are cleared or
the instrument was not calibrated in the pH range.
*
" on the right side of the buffer value.
31

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hi 98184Hi 98181Hi 98185Hi 98182Hi 98183

Table of Contents