Hanna Instruments HI9811-5 Instruction Manual
Hanna Instruments HI9811-5 Instruction Manual

Hanna Instruments HI9811-5 Instruction Manual

Portable ph/ec/tds/c meters
Hide thumbs Also See for HI9811-5:

Advertisement

Quick Links

Instruction Manual
HI9811-5
Portable
pH/EC/TDS/°C Meters
www.hannainst.com
1

Advertisement

Table of Contents
loading

Summary of Contents for Hanna Instruments HI9811-5

  • Page 1 Instruction Manual HI9811-5 Portable pH/EC/TDS/°C Meters www.hannainst.com...
  • Page 2: Table Of Contents

    If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid.
  • Page 3: Preliminary Examination

    Any defective item must be returned in the original packing together with the supplied accessories. GENERAL DESCRIPTION HI9811-5 is the complete, versatile, with a large LCD and splash- proof portable combination meter designed with the utmost precision and simplicity. The instrument provides measurements for pH, EC and TDS ranges, which are easily selectable through the keypad on the front panel.
  • Page 4: Functional Description

    FUNCTIONAL DESCRIPTION 1) 8-pin DIN connector for probe 2) Liquid Crystal Display 3) pH range selection key 4) µS/cm (EC) range selection key 5) mg/L (TDS) selection key 6) pH offset calibration knob 7) EC/TDS calibration knob 8) °C (Temperature) selection key 9) On/Off key...
  • Page 5: Specifications

    SPECIFICATIONS 0.0 to 14.0 pH 0 to 6000 µS/cm Range 0 to 3000 mg/L (ppm) 0.0 to 70.0 °C 0.1 pH 10 µS/cm Resolution 10 mg/L (ppm) 0.1 °C ±0.1 pH Accuracy ±2% f.s. µS/cm (@ 20 °C/68 °F) ±2% f.s. mg/L (ppm) ±0.5 °C ±0.1 pH ±2% f.s.
  • Page 6: Operational Guide

    OPERATIONAL GUIDE INITIAL PREPARATION Each meter is supplied complete with a 9V battery. Remove the battery compartment cover on the back of the meter and install the battery while observing its polarity. Connect the probe to the DIN socket on the top of the meter by aligning the pins with the socket and pushing in the plug.
  • Page 7 • Select the appropriate measurement range (EC or TDS). • Wait a couple of minutes for the temperature sensor to reach thermal equilibrium. The display will then show the measurement automatically temperature compensated for temperature with the appropriate indication among the following: “µS”...
  • Page 8: Ph Calibration

    pH CALIBRATION For greatest accuracy, frequent calibration of the instrument is recom- mended. The instrument should be recalibrated for pH: a) Whenever the electrode is replaced. b) At least once a month. c) After testing aggressive chemicals. d) Where extreme accuracy is required. PREPARATION Pour small quantities of pH7.01 (HI7007) or pH4.01 (HI7004) or pH10.01 (HI7010) solution into a clean beaker.
  • Page 9 Notes: • The probe should be submerged approximately 4 cm (1½”) into the solution. The thermometer has to be close to the probe. • If turning the knob the needed value can not be reached, clean the probe (see the “Probe Maintenance” section). If also after the probe cleaning the value can not be reached, replace the probe.
  • Page 10: Ph Values At Various Temperatures

    pH VALUES AT VARIOUS TEMPERATURES For temperature compensation during calibration, please refer to the following chart. TEMP pH VALUES °C °F 4.01 6.86 7.01 9.18 10.01 4.01 6.98 7.13 9.46 10.32 4.00 6.95 7.10 9.39 10.24 4.00 6.92 7.07 9.33 10.18 4.00 6.90...
  • Page 11: Ec/Tds Calibration

    The TDS value in aqueous solutions is directly proportional to the con- ductivity. The ratio between the two parameters depends on the solution. HI9811-5 model has a fixed conversion factor set to 0.5. This means that 1 µS/cm is equal to 0.5 mg/L of TDS.
  • Page 12: Battery Replacement

    BATTERY REPLACEMENT The meters is powered by a 9V battery that is located on the rear of the instrument. When battery symbol is empty a low battery condition is indicated. When the low battery indication appears, only a few hours of working time remains.
  • Page 13: Probe Maintenance

    For better accuracy in measurements and to ensure a good performance of the probe, a frequent cleaning is recommended. For this purpose, soak it in Hanna Instruments HI700661 Cleaning Solution for 5 minutes. Notes: • For particular dirt (as for example protein, oil or grease) see the “Accessories”...
  • Page 14: Accessories

    ACCESSORIES PROBES HI1285-0 Combination, amplified pH/EC/TDS probe with built-in temperature sensor, 8-pin DIN connector and 1 m (3.3’) cable HI1285-5 Combination, amplified pH/EC/TDS/temperature probe with built-in temperature sensor, 8-pin DIN connector and 1 m (3.3’) cable pH BUFFER SOLUTIONS HI7004L pH4.01 buffer solution, 500 mL bottle HI7006L pH6.86 buffer solution, 500 mL bottle...
  • Page 15 For yours and the meter’s safety do not use or store the meter in hazardous environments. Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice.
  • Page 16 World Headquarters Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA www.hannainst.com Local Office Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA Phone: 800.426.6287 Fax: 401.765.7575 e-mail: tech@hannainst.com MAN9811 06/18...