Citizen CH-302B Instruction Manual

Citizen CH-302B Instruction Manual

Digital blood pressure monitor

Advertisement

TM
INSTRUCTION MANUAL
FOR
DIGITAL BLOOD PRESSURE MONITOR
REF CH-302B

IDENTIFICATION OF PARTS

Exhaust button
Bulb
Air-release system
Systolic blood pressure display section
Diastolic blood pressure/Pulse display section
Bulb hose
Cuff
Cuff hose
Hose connector
Power switch
Battery cover
- Eng 1 -

Advertisement

Table of Contents
loading

Summary of Contents for Citizen CH-302B

  • Page 1: Identification Of Parts

    INSTRUCTION MANUAL DIGITAL BLOOD PRESSURE MONITOR REF CH-302B IDENTIFICATION OF PARTS Exhaust button Bulb Air-release system Systolic blood pressure display section Diastolic blood pressure/Pulse display section Bulb hose Cuff Cuff hose Hose connector Power switch Battery cover - Eng 1 -...
  • Page 2: Specifications

    SPECIFICATIONS Measuring method: Oscillometric system Indication: Digital display, Height: 10mm Measuring range: Pressure; 0~280mmHg, Pulse; 40~200 beat/min. Accuracy: Pressure; 3mmHg, Pulse; 5% Inflation: Manual inflation by bulb Deflation: Automatic constant speed deflation system Power source: One battery (6F22, 6LR61) Battery life: Approx.
  • Page 3: General Remarks

    CLEANING METHOD OF CUFF After cleaning the cuff with a neutral detergent, lightly rinse it a washing machine will cause air leakage. Also, be careful not with water and let it dry in the air. Hard rubbing or washing in to get water into the air hose.
  • Page 4: Precaution For Use And Maintenance

    • Measure blood pressure with a relaxed and quiet posture. • Do NOT keep on measuring so many times, as your arm will Keep the center of the cuff at the level of your heart, and do have built up congestion of blood, and you will not get a valid NOT move your arm or talk during measurement.
  • Page 5: Preparation For Measurements

    Protect the nature environment: Manufacturer: Please help protect natural environment by respecting national Name: Premier Precision Ltd. and/or local recycling regulations when disposing of the bat- Address: Block E, G-H, 6 floor, teries and the products at the end of their useful lives. Phase 2, Yip Fat Factory Building, 73-75 Hoi Yuen Road, Kwun Tong, Kowloon, Hong Kong...
  • Page 6: Attach The Cuff

    3. Attach the cuff The cuff hose should be placed on the inside of your arm as figures. The cuff should be wound so that one finger can be inserted Wind the arm cuff around the left upper arm tightly where the under the cuff.
  • Page 7 MEASUREMENTS PROCEDURES 1. Power on SYSTOLIC mmHg DIASTOLIC mmHg If there is some air re- SYSTOLIC mmHg DIASTOLIC mmHg maining in the cuff, release Touch the power switch air with the exhaust button for power on. until the “ 0 ” appears. The PULSE/min All display marks will be buzzer beeps 3 times.
  • Page 8: Power Off

    When pressurization is 4. Press the exhaust button SYSTOLIC mmHg DIASTOLIC mmHg completed, the automatic SYSTOLIC mmHg DIASTOLIC mmHg air-release system will Press the exhaust button gradually reduce cuff to completely release the pressure, and Measuring PULSE/min air within the cuff. Mark “Q”...
  • Page 9: Description Of Display Marks

    DESCRIPTION OF DISPLAY MARKS DISPLAY MARK CONDITION/CAUSE CORRECTIVE ACTION SYSTOLIC mmHg DIASTOLIC mmHg The cuff was not inflated to more than Raise the set pressure value by 30-40 mmHg above your systolic pressure. higher level PULSE/min SYSTOLIC mmHg DIASTOLIC mmHg The cuff was inflated to more than 320 mmHg.
  • Page 10 CITIZEN is a registered trademark of Citizen Watch Co., Japan. Consumer Electronics Division CBM Bldg. 5F, 5-68-10, Nakano, Nakano-ku, Tokyo 164-0001, Japan Phone: (03)5345-7430 Fax: (03)5345-7431 E-mail: info-con@jcbm.co.jp...