Possible error messages.
The following messages can be shown on the test tool display:
Proceed as follows:
The calibration date and number will not be updated. You must continue with the Final
Calibration!
5-18
WARNING: Calibration data not valid.
Save data and exit maintenance mode?
If you did the WarmingUp and Pre-Calibration successfully (section 5.5), and you
want to store the Pre-Calibration data before continuing with the Final Calibration:
⇒ Press
.
YES
When turning the test tool off and on again, it will show the message:
The instrument needs calibration.
Please contact your service center.
To return to the Maintenance mode, if you want to repeat the complete calibration:
⇒ Press
.
NO
Now press
until the display shows
calibrate the test tool, starting at section 5.5.
If you want to exit and maintain the old calibration data:
⇒ Turn the test tool off.
WarmingUp (CL 0200):IDLE