Fluke 19xB-19xC-2x5C
Service Manual
5. Using
6. Set the calibrator format and marker line number to :
7. Observe the trace, and check to see if the test tool triggers on the negative pulse
8. Select the following test tool setup:
9. Set the calibrator video trigger output signal to -100%
10. Using
11. Set the calibrator format and marker line number to :
12. Observe the trace, and check to see if the test tool triggers on the positive pulse
4-26
select test tool line number:
⇒ 310 for PAL, PALplus or SECAM
⇒ 262 for NTSC
⇒ PAL 310 (odd), for PAL and PALplus
⇒ SECAM 310 (odd), for SECAM
⇒ NTSC 262 odd, for NTSC.
before the marker.
Press
to open the Trigger Options menu.
Choose
VIDEO on A...
Polarity:
NEGATIVE |
select line number 310 (PAL, PALplus or SECAM) or 262 (NTSC)
⇒ PAL 310 (odd), for PAL and PALplus
⇒ SECAM 310 (odd), for SECAM
⇒ NTSC 262 odd, for NTSC.
before the marker.
Figure 4-17. SC600 Marker Pulse
, then from the shown opened menu choose
( or
PAL
NTSC
)
PALplus
SECAM
video-sc600.bmp