Emissivity - Table And Unknown Value - Fluke 574 User Manual

Precision infrared thermometer
Hide thumbs Also See for 574:
Table of Contents

Advertisement

Emissivity - Table and Unknown Value

T
TABLE OF VALUES
TABLE OF VALUES
To choose the emissivity
of a material, press
EMISS (P). The display
shows a material name
(7), an emissivity value,
and the calculated
temperature value (5). To
choose another material,
use the Up and Down
keys (C). Press ENTER
(D) to activate this setting.
UNKNOWN VALUE
T
To adjust the unit'
To adjust the unit'
emissivity value for a
material with unknown
emissivity, plug in the
probe.
Pull the unit's trigger.
Place the measuring
tip of the probe on the
area to be measured.
Wait for the reading to
stabilize.
Note the indicated probe temperature reading.
Release the trigger. Pull the trigger again. Measure
the same area using infrared measurement. Press
the emissivity button (P). Use the Up and Down keys
(C) to select the material name "Free" which will be
shown in the display (7). Press the emissivity button
(P) again until the emissivity sign (5) flashes. Use
the arrow keys (C) to change the emissivity value
until the temperature matches the probe's reading.
s
P
7
574
5
C
D
15