Ma Dc Source Adjust Procedure; Volts Dc Source Adjust Procedure (773 Only) - Fluke 772 Calibration Manual

Milliamp process clamp meter
Hide thumbs Also See for 772:
Table of Contents

Advertisement

772/773
Calibration Manual

mA DC Source Adjust Procedure

To adjust the mA Source function, use the Calibrator to apply the necessary Meter input
and do the steps in Table 9.
Step
1
2
4

Volts DC Source Adjust Procedure (773 Only)

To adjust the Volt Source function, use the Calibrator to apply the necessary Meter input
and do the steps in Table 10.
Step
1
2
4
14
Table 9. mA Source Adjustment Procedure
Meter LCD display
4.00 mA
20.00 mA
Save
Table 10. Volt Source Adjustment Procedure
Meter LCD display
0.00 V
10.00 V
Save
Action
Adjust until Meter output is 4.00 mA, push H
Adjust until Meter output is 20.00 mA, push H
Push H
Action
Adjust until Meter output is 0.00 V, push H
Adjust until Meter output is 10.00 V, push H
Push H