IMPORTANT CONTROL SOLUTION TEST INFORMATION
■
Use only OneTouch
■
Check the expiration date of the control solution. Do not use if expired.
■
Record the discard date on the control solution vial. The discard date for
control solution is three months after first opening the vial. Discard the
vial after this date.
■
The OneTouch
®
for OneTouch
®
range for your blood glucose level.
Store the control solution at room temperature below 86˚F (30˚C). Do not
■
refrigerate.
Control solution should not be used as a cleaning solution for the meter.
■
Warning
For external use only. Do not swallow or inject control solution or put
■
control solution drops into your eyes.
Normal Control Solution.
®
Normal Control Range printed on the test strip vial is
Normal Control Solution only. It is not a recommended
22