Cautions Regarding Temperature / Humidity Conditions for CA-2000
It is recommended that the instrument be used and stored under standard conditions (T
humidity: 40 %), and that areas subject to high temperature and/or humidity be avoided. In addition in order to
maintain the measurement accuracy of this instrument, it is recommenced that it be inspected regularly about once
a year
. For details on having the instrument inspected, please contact the nearest Konica Minolta Sensing
authorized service facility
Even if the product is used within the specified operating temperature/humidity range or stored within the specified
storage temperature/humidity range, the displayed values may be af
If the instrument is left under the following high-temperature conditions for a long period of time, the displayed
values may change as follows:
T
emperature: 30
Accuracy: Luminance:
Inter-point error: Luminance:
T
emperature: 35
Accuracy: Luminance:
Inter-point error: Luminance:
These dif
ferences in display values are due to the instrument materials and/or components being affected by the
temperature and humidity conditions of long-term use or storage. In particular
temperature or humidity
.
changes, the accumulated ef
Printed in Japan
.
C; Relative humidity:70 %; Period under these conditions: 720 hours (30 days)
0.4 %; Chromaticity:
C; Relative humidity:55 %; Period under these conditions: 336 hours (14 days)
1 %; Chromaticity:
Although measures have been taken to improve resistance to temperature/humidity
fect of long-term use or storage may affect the displayed values.
fected by long-term conditions of use or storage.
0.003
0.2 %; Chromaticity:
0.0003
0.006
0.5 %; Chromaticity:
0.001
, optical filters are easily af
emperature: 23 C; Relative
fected by
AHDAPX ( 2 )