Recommended separation distances between portable or mobile RF communication equipment
and the RadiForce Series
The RadiForce series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the RadiForce series can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the RadiForce series.
Immunity to proximity fields from following RF wireless communication equipments has been confirmed:
Test fre-
Bandwidth
a)
quency
(MHz)
(MHz)
385
380 - 390
TETRA 400
450
430 - 470
GMRS 460,
FRS 460
710
704 - 787
LTE Band 13, 17
745
780
810
800 - 960
GSM 800 / 900,
TETRA 800,
870
iDEN 820
930
CDMA 850,
LTE Band 5
1720
1700 - 1990 GSM 1800;
CDMA 1900;
1845
GSM 1900;
1970
DECT;
LTE Band 1, 3, 4,
25;
UMTS
2450
2400 - 2570 Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
5240
5100 - 5800 WLAN 802.11 a/n Pulse modulation
5500
5785
a) For some services, only the uplink frequencies are included.
b) carrier are modulated using a 50 % duty cycle square wave signal.
The RadiForce series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. For other por-
table and mobile RF communication equipments (transmitters), minimum distance between portable and mobile RF communications equipment
(transmitters) and the RadiForce series as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance "d" in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where "P" is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1
At 80 MHz and 800 MHz, the separation distance for a higher frequency range applies.
Note 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Service
a)
Modulation
Pulse modulation
18 Hz
FM
±5 kHz deviation
1 kHz sine
Pulse modulation
217 Hz
Pulse modulation
18 Hz
Pulse modulation
217 Hz
Pulse modulation
217 Hz
217 Hz
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
Minimum
Maximum
separation
b)
power
distance
test level (V/m)
(W)
(m)
b)
1.8
0.3
27
2
0.3
28
b)
0.2
0.3
9
b)
2
0.3
28
b)
2
0.3
28
b)
2
0.3
28
b)
0.2
0.3
9
(m)
80 MHz to 800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
Compliance
IEC / EN60601
level
(V/m)
27
28
9
28
28
28
9
800 MHz to 2.7 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23
Appendix