Electromagnetic immunity
The RadiForce series has been tested at the following compliance levels according to the testing requirements for profes-
sional healthcare facility environments defined in IEC / EN60601-1-2.
Customers and users of the RadiForce series must ensure that the RadiForce series is used in the following environments:
Immunity test
Test level for professional
healthcare facility
environments
Electrostatic dis-
±8 kV contact discharge
charge (ESD)
±15 kV air discharge
IEC / EN61000-4-2
Electrical fast tran-
±2 kV power lines
sients / bursts
±1 kV input / output lines
IEC / EN61000-4-4
Surges
±1 kV line to line
IEC / EN61000-4-5
±2 kV line to ground
Voltage dips, short
0 % U
T
interruptions and
0.5 cycles and 1 cycle
voltage variations on
70 % U
power supply input
25 cycles
lines
0 % U
T
IEC / EN61000-4-11
5 sec
Power frequency
30 A/m
magnetic fields
(50 / 60 Hz)
IEC / EN61000-4-8
Compliance level
±8 kV contact discharge
±15 kV air discharge
±2 kV power lines
±1 kV input / output lines
±1 kV line to line
±2 kV line to ground
(100 % dip in U
)
0 % U
(100 % dip in U
T
T
0.5 cycles and 1 cycle
(30 % dip in U
)
70 % U
(30 % dip in U
T
T
T
25 cycles
(100 % dip in U
)
0 % U
(100 % dip in U
T
T
5 sec
30 A/m
Electromagnetic environment - Guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
)
Mains power quality should be that of a typi-
T
cal commercial or hospital environment. If the
)
user of the RadiForce series requires continued
T
operation during power mains interruptions, it
)
is recommended that the RadiForce series be
T
powered from an uninterruptible power supply or
a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
The product should be kept at least 15 cm away from
the source of power frequency magnetic fields during
use.