1577 Accuracy
Range
600.0 mV
0.1 mV
6.000 V
0.001 V
60.00 V
0.01 V
600.0 V
0.1 V
1000 V
1 V
AC Conversion ................................................... Inputs are ac-coupled and calibrated to the rms value of sine wave
Input Impedance ................................................. 10 MΩ (nominal), <100 pF, ac-coupled
Common Mode Rejection Ratio
(1 kΩ unbalanced) .............................................. >60 dB at dc, 50 or 60 Hz
DC Voltage Measurement
Range
6.000 V dc
0.001 V
60.00 V dc
0.01 V
600.0 V dc
0.1 V
1000 V dc
1 V
Accuracies apply to ± 100% of range.
[1]
Input Impedance ................................................. 10 MΩ (nominal), <100 pF
Normal Mode Rejection Ratio ............................. >60 dB @ 50 Hz or 60 Hz
Common Mode Rejection Ratio .......................... >120 dB @ dc, 50 Hz or 60 Hz (1 kΩ unbalance)
36
+7 (495) 109-02-08 sales@bbrc.ru www.bbrc.ru
Resolution
input. Conversions are true-rms responding and specified from 5 % to
100 % of range. Input signal crest factor can be up to 3 at up to 500 V,
decreasing linearly to crest factor ≤1.5 at 1000 V. For non-sinusoidal
waveforms add ±(2 % reading + 2 % FS) typical, for a crest factor up
to 3.
Accuracy 1587 and 1587T
Resolution
±(% of Rdg + Counts)
±(0.09 % + 2)
±(0.09 % + 2)
±(0.09 % + 2)
±(0.09 % + 2)
50 Hz to 60 Hz
±(% of Rdg + Counts)
±(2 % + 3)
±(2 % + 3)
±(2 % + 3)
±(2 % + 3)
±(2 % + 3)
[1]
[1]
Accuracy 1577
±(% of Rdg + Counts)
±(0.2 % + 2)
±(0.2 % + 2)
±(0.2 % + 2)
±(0.2 % + 2)