Download Print this page

Fluke 8024B Instruction Manual page 74

Hide thumbs Also See for 8024B:

Advertisement

8024B
DMM
CALIBRATOR
Figure 4-4. General Equipment Connection - Current
Table 4-4. Current Test
"POSITION.
per
INPUT
STEP
POSITION
ac/bc | RANGE|
LEVEL
DC
+190 mA dc
200 mA | _190 mA de
2000 mA}
+1.9Adc
+1885 to +1915
}
0.000 to 0.002
2mA
;
20 mA
{19 mA acrms .
18.69 to 19.31
Ac
200 mA
{190 mA ac rm [100 He
186.9 to 193.1
2000 mA]
1.9A ac rms
| 100 He |
1869 to 1931
DISPLAY
LIMITS
+1.885 to +1.915
:
+18.85 to +19.15
+188.5 to +191.5
—188.5 to —191.5
4-10

Advertisement