Fluke 8024B Instruction Manual page 37

Hide thumbs Also See for 8024B:

Advertisement

8024B
2-20
9.
Depress the © function switch and the AC/DC switch. The A up arrow will
appear in the display. Sequentially depress each of the six range switches. The
display will indicate an overrange condition and the decimal point will change
position.
10.
Touch the red probe tip to the COMMON
input terminal, the audible tone
will sound and the Wdown arrow will appear in the display. Sequentially depress
each of the grey range switches. The display should read zero at each range setting.
Lead resistance may be sufficient to cause a one or two tenths (0.1 or 0.29)
indication on the 2000 range. Release the AC/ DC switch to silence the audio tone.
11.
Touch the red probe tip to the mA ~°C input connector and press the 2000
switch. The display should read 99.0 to 101.0.
12.
Press the 2 kO switch. The display should read .099 to .101. Remove the probe
from the mA -°C input connector.
13.
Simultaneously depress the two range switches to select the 200 nS range. The
display should read 00.0 to 01.0 (minimum conductance, maximum resistance).
14.
Touch the red probe tip to the COMMON
input terminal. An overrange
indication should be displayed since conductance is the reciprocal of resistance.
15.
Connect the black test lead to the COMMON
input connector.
16.
Depress both the AC/DC switch and the 750V ac range switch. Set the
function switch to the voltage (out) position. (Use 750V range for 230V line.)
WARNING
THE LOCAL LINE VOLTAGE IS MEASURED IN THE FOLLOWING STEP. BE
CAREFUL
NOT TO TOUCH
THE PROBE TIPS WITH
FINGERS,
OR TO
ALLOW THE PROBE TIPS TO CONTACT EACH OTHER.