Download Print this page

Fluke 8024B Instruction Manual page 66

Hide thumbs Also See for 8024B:

Advertisement

8024B
INSTRUMENT
TYPE
DMM
Calibrator
DMM
Temperature
Reference
Monitor
Thermocouple
Accessory
Vacuum
Insulated Bottle
2 Hole Cork
Pulse
Generator
4-2
Table 4-1. Test Equipment Required
REQUIRED CHARACTERISTICS
AC VOLTS:
Voltage Range: OV to 200V to
750V
Accuracy Required: +.1%,
+0.35%, +0.1%
Frequency: 5 kHz to 2 kHz to 1
kHz to 100 Hz, +0.25%
DC VOLTS
Voltage Range: 0 to 1000V
Accuracy: +0.025%
AC CURRENT:
Current Range: 0 to 1900 mA
Frequency Range: 100 Hz to
1 kHz
Accuracy: +0.1%
DC CURRENT:
Current Range: 0 to 1900 mA
Accuracy: +0.1%
RESISTANCE:
Value: 10092
Accuracy: +0.05%
Value:
1kQ, 10 kQ, 100 kQ,
1000 kQ
Accuracy: +0.025%
Value: 10 MQ.
Accuracy: +0.5%
0 to 12V dc +.1%
0.4°C resolution at 0°C
8024B compatible
1 quart capacity minimum
To fit mouth of Vacuum Insulated
Bottle
Can generate 25 psec pulses at
SV p-p
RECOMMENDED
EQUIPMENT
John Fluke
Model! 5100B
John Fluke
Model 8020B
PRINCO ASTM
56C
John Fluke
Model Y8102
or Y8103
Thermos
Hewlett Packard
Model 8003A

Advertisement

loading