Chapter 6
Digital I/O

Digital Waveform Acquisition

Figure 6-2 summarizes all of the timing options provided by the digital input timing engine.
100 MHz Timebase
DSTAR <A..B>
PFI, RTSI
Revamp AO timing options with DI timebase and clocks
PXI_STAR
Analog Comparison
Event
20 MHz Timebase
100 kHz Timebase
PXI_CLK10
You can acquire digital waveforms on the Port 0 DIO lines. The DI waveform acquisition FIFO
stores the digital samples. X Series devices have a DMA controller dedicated to moving data
from the DI waveform acquisition FIFO to system memory. The DAQ device samples the DIO
lines on each rising or falling edge of a clock signal, DI Sample Clock.
You can configure each DIO line to be an output, a static input, or a digital waveform acquisition
input.
X Series devices feature the following digital input timing signals:

DI Sample Clock Signal

DI Sample Clock Timebase Signal
DI Start Trigger Signal
DI Reference Trigger Signal
DI Pause Trigger Signal
*
Signals with an
support digital filtering. Refer to the
more information.
DI Sample Clock Signal
The device uses the DI Sample Clock (di/SampleClock) signal to sample the Port 0 terminals
and store the result in the DI waveform acquisition FIFO.
You can specify an internal or external source for DI Sample Clock. You can also specify
whether the measurement sample begins on the rising edge or falling edge of DI Sample Clock.
6-4 | ni.com
Figure 6-2. Digital Input Timing Options
Analog Comparison Event
DI Sample Clock
Timebase
*
*
*
*
DSTAR <A..B>
PFI, RTSI
PXI_STAR
Ctr n Internal Output
Programmable
Clock
Divider
PFI Filters
section of Chapter 8, PFI, for
DI Sample Clock