Invacare® M6, M9, D Cylinder with Continuous Flow Regulator
The cylinder assembly MUST be emptied of gas prior to shipping by
courier or postal service. Refer to Emptying the Cylinder Assembly.
Because the cylinder contains magnetic, ferrous material, it is not
suitable to be in an MRI room.
Be sure all connections are tight and leak free. DO NOT use a liquid
leak detector to test for leaks. Leaks are indicated by loss of pressure
when the cylinder assembly is not in use and by hissing sounds. This
product is equipped with a relief valve. If you hear a loud hissing or
popping sound coming from the product, discontinue use of product,
turn the flow dial to the "0" setting, and contact your dealer/Home
Care equipment provider.
ALWAYS confirm prescribed dose before administering to patient
and monitor on a frequent basis.
NO OXYGEN is delivered when the regulator flow dial setting is "0".
DO NOT use a humidifier with this product.
Keep cylinder in a well ventilated place.
This product is designed to operate with a standard or high flow
single lumen, adult oxygen nasal cannula with a maximum length of
2.1 m (7 ft).
The cannula is for single patient use only.
Cannula MUST be installed on outlet connection prior to use.
Refer to the HomeFill User Manual for additional warnings, cautions
and safety instructions.