Analog Input Ground-Reference Settings - National Instruments PXIe-6355 User Manual

Hide thumbs Also See for PXIe-6355:
Table of Contents

Advertisement

Chapter 4
Analog Input

Analog Input Ground-Reference Settings

MIO X Series devices support the following analog input ground-reference settings:
Differential mode—In DIFF mode, the MIO X Series device measures the difference in
voltage between two AI signals.
Referenced single-ended mode—In RSE mode, the MIO X Series device measures the
voltage of an AI signal relative to AI GND.
Non-referenced single-ended mode—In NRSE mode, the MIO X Series device measures
the voltage of an AI signal relative to one of the AI SENSE inputs specific for that channel.
The AI ground-reference setting determines how you should connect your AI signals to the MIO
X Series device. Refer to the
Ground-reference settings are programmed on a per-channel basis. For example, you might
configure the device to scan 12 channels—four differentially-configured channels and eight
single-ended channels.
MIO X Series devices implement the different analog input ground-reference settings by routing
different signals to the NI-PGIA. The NI-PGIA is a differential amplifier. That is, the NI-PGIA
amplifies (or attenuates) the difference in voltage between its two inputs. The NI-PGIA drives
the ADC with this amplified voltage. The amount of amplification (the gain), is determined by
the analog input range, as shown in Figure 4-2.
4-4 | ni.com
Connecting Analog Input Signals
Figure 4-2. MIO X Series Device NI-PGIA
V
in+
PGIA
V
in–
V
m
V
= [V
– V
] × Gain
m
in+
in–
section for more information.
+
Measured
Voltage

Advertisement

Table of Contents
loading

Table of Contents