17.1 IO-Link process data structure

Process data length: 9 bytes
Byte 
Data 
no. 
0 ‐ 2  Flow 
3 ‐ 6  Volume 
7 ‐ 8  Temperature 
reserved 
reserved 
Status OUT1 
Status OUT2 
page 50
Bit count 
Format 
24 Bit 
IntegerT  1/1000  +/‐8388,608 
32 Bit 
FloatT 
12 Bit 
IntegerT  1/10 
1 Bit 
BooleanT    
1 Bit 
BooleanT    
1 Bit 
BooleanT    
1 Bit 
BooleanT    
Factor 
Range 
‐45
  
+/‐1,4*10
 ... +/‐3,4*10
+/‐204,8 
  
  
  
  
Value 
L/min 
+38
 
°C 
  
  
  
  
A5E48762204-AB