Process conditions

9 Operating conditions

9.1 Process conditions
The pressure and temperature ratings presented in this section represent the
design values for the devices. For individual applications (e.g. marine
applications with option S2x) further limitations may apply according to the
respective applicable regulations. For details see chapter Application and
industry related standards [} 185] under the heading Marine approvals.
In this chapter, all values related to pressure are gauge pressure values.

9.1.1 Process fluid temperature range

Allowed process fluid and ambient temperature ranges in hazardous areas
depend on classifications defined by applications, refer to Temperature
specification in hazardous areas [} 54].
For SITRANS FC6x0 the following process fluid temperature ranges are available:
Temperature
range
Standard
Low
Medium
High
1)
With process connection type G2 and G6 temperature range is: -10 – 140 °C (14 – 284
°F)
2)
With SITRANS FC HART firmware rev.4 or later. For details please contact your local
Siemens sales organization.
7 M E 4 4 x
-
7
5 6
Order code
position 12
1)
1, 6
2)
2
3, 7
4, 8
-
8 9 10 11
12
13 14 15 16
Process fluid
Design type
temperature
in °C
(°F)
-50 – 150
Compact type
(-58 – 302)
-70 – 150
(-94 – 302)
-196 – 150
(-321 – 302)
Remote type
-70 – 230
(-94 – 446)
0 – 350
(32 – 662)
A5E52748375-AA, 2023/09
FC6x0
Operating conditions
-
Z
Order code
position 14
A, B
C, D, E, F, G,
H
D, F, H
39 / 194