As an option, one of the switching inputs can be configured to reset the
min/max memory and to define the smoothing filter setting.
target object in the pyrometer's field of view
Automatic temperature detection (ATD) function*
This function serves to automatically detect the temperature during
manufacturing processes with discontinuous or intermittent material flow,
for example molten metal casting at foundries. First, define the sampling
time and the upper and lower temperature limits. These limits or thresh-
olds define the temperature range within which you wish to detect the
temperature. It is also possible to determine the average temperature
value over the course of multiple measuring cycles.
The start of a measuring cycle is determined automatically and is de-
pendent on the following variables.
Limit 1 (LI. 1)
Limit 2 (LI. 2)
T.DEL
Time Delay (
When the conditions are fulfilled, the sampling time can begin. (
temp. reading with double max. buffer
holdtime
temp. measurement without double max. buffer
Before beginning the measurement, the
temperature reading must have been lower
than Limit 1 at least once.
If Autoreset (
be ignored
Limit 2 must be exceeded at least for the
duration of the time delay (
See above
):
19
Operating Manual MPA
A.RST= ON
) the limit 1 will
T.DEL
time
).
T.ACT
).