Appendix B: Time-out Settings

The following procedure provides the instructions on how to set the
time- out value in the Agilent IO Libraries.
1 To launch your Agilent Connection Expert, go to Start > All Programs >
2 On the Instrument I/O on this PC panel, select your device in the USB
3 Select Send Commands To This Instrument and the Agilent Interactive IO
4 Go to Interactive IO > Options and the Options dialog box will appear.
5 Configure the time- out value based on the table provided below.

Table 4-2 Time-out settings

Points
1 to 10
11 of 50
51 to 100
101 to 500
501 to 1000
1001 to 4096
N O T E
60
Agilent IO Libraries Suite > Agilent Connection Expert.
expanded list and right- click.
window will appear.
• Expected time, t = Points
• The Points is set using the SENSe:SWEep:POINts command and
the Tint is set using the SENSe:SWEep:TINTerval command.
Tint (ms)
1 to 100
101 to 32767
1 to 50
51 to 32767
1 to 10
11 to 32767
1 to 5
6 to 500
501 to 32767
1 to 500
501 to 32767
1 to 500
501 to 32767
×
Tint
Time-out range (ms)
Std IO 5000
Assign 110% of t
Std IO 5000
Assign 120% of t
Std IO 5000
Assign 120% of t
Std IO 5000
Assign 130% of t
Assign 110% of t
Assign 310% of t
Assign 120% of t
Assign 310% of t
Assign 110% of t
U2722A/U2723A User's Guide