Omron BP7250CAN Instruction Manual page 36

Upper arm blood pressure monitor
Table of Contents

Advertisement

Display/Problem
appears or a
measurement
cannot be complete
after the arm cuff
inflates.
appears
appears
Possible Cause
You move or talk during
a measurement and the
arm cuff does not inflate
sufficiently.
The systolic pressure is
above 210 mmHg and a
measurement cannot be
taken.
The arm cuff is inflated
exceeding the maximum
allowable pressure.
You move or talk during
a measurement.
Vibrations disrupt a
measurement.
36
Solution
Remain still and do not
talk during a
measurement. If "E2"
appears repeatedly,
inflate the arm cuff
manually until the
systolic pressure is 30 to
40 mmHg above your
previous readings. Refer
to sub-section 3.1.
Inflate the arm cuff
manually until the
systolic pressure is 30 to
40 mmHg above your
previous readings. Refer
to sub-section 3.1.
Do not touch the arm cuff
and/or bend the air tube
while taking a
measurement.
If inflating the arm cuff
manually, refer to the
end of sub-section 3.1.
Remain still and do not
talk during a
measurement.

Advertisement

Table of Contents
loading

This manual is also suitable for:

5 series