Setting of Users' Ranges
It is available to individually set the suitable value responds to each sample in seven users' ranges. Angle and
equilibrium of a relation function are editable for each range.
Parallel change (equilibrium change)
1. After turning on the device, Press + or − key to choose the
range to be edited.
2. Press + and − keys simultaneously.
The character at upper right changes to "H" and the current
equilibrium is shown. The default setting is "0.0" in all ranges.
3. Press + or − key to modify the value, available between −99
and 99.9. It can vary in steps of 0.1 between −19.9 and 99.9.
The value increases or decreases continuously by a long press.
In the right case, set it to −3.0.
4. Press AVE/OFF key to complete. If a measured value was
displayed before this setting, a recalculated value is displayed.
Otherwise, "L" is displayed.
Angle change
1. After turning on the device, Press + or − key to choose the
range to be edited.
2. Press + and − keys simultaneously, and then press START key.
The character at upper right changes to "A" and the current
angle is shown. The default setting is "100" in all ranges.
3. Press + or − key to modify the value, available between 0 and
999. The value increases or decreases continuously by a long
press.
In the right case, set it to 66 following the equation A*.
4. Press AVE/OFF key to complete.
Specifications
Name:
Measuring method :
Measuring range:
Operating ambient
Resolution:
Power requirement:
Dimensions:
Weight:
Accessories:
Warranty Policy
Our products are warranted to be free from defects in materials and workmanship for a period of one year from date
of delivery. If repair is necessary and has not been the result of misuse, force majeure, or transportation arranged by
yourself within the one-year period, please return the products on freight prepaid basis. Correction of the defect will
be made without charges. We alone will determine if the product problem is due to deviations or customer misuse.
Out-of-warranty products will be repaired for a fee.
Before returning, request for our acknowledgement first.
For details, contact us or a retailer from which you have purchased.
SATO KEIRYOKI MFG. CO., LTD.
3-4, Kanda-kajicho, Chiyoda-ku, Tokyo 101-0045, Japan
https://www.sksato.co.jp/en/
Moisture Meter for Compost Model SK-970A
Electric resistance measurement
23.0 to 80.0% in the standard range
5 to 40°C without condensing
0.1%
Two AA zinc-carbon batteries
Main body: 70 (W) x 155 (H) x 25 (D) mm
Cable: 0.8 m long
Sensor: 10 mm dia. x 800 mm long
447 g including batteries
Two AA zinc-carbon batteries, hand strap
Pattern 1: When all the measured
Pattern 1: When all the measured
values are 3% higher
values are 3% higher
Measured
value
Pattern 2: When 60%-water compost is
detected correctly whereas 70%-water
one is measured 5% higher
Measured
value
Difference of known values
A* =
Difference of measured values
2
Function
before
setting
Scatter
plots
Set to
−3%
Known moisture content
Function
after setting
Set to
66%*
Plots
of 10%
Plots
of 20%
Known moisture content
70%−60%
=
= 66%