To calibrate the thermohygrometer, use a temperature/humidity generating chamber to
produce reference temperature and reference humidity. A chamber capable of achieving a
temperature repeatability within ±0.1°C, humidity repeatability within ±1.0%, and humidity
variability between 0 and 90% is required.
Equipment
110TRH
Temperature/humidity generating chamber
Reference thermometer
Torque screwdriver
Phillips head screwdriver
Flat head screwdriver
1. Remove the six screws from the device as shown.
2. Remove the front section.
3. Insert four batteries into the case in the directions indicated.
4. Press and hold the POWER key for more than one second to activate the device.
5. Set the sensor in the temperature/humidity generating chamber and allow it to adjust to
the temperature and humidity.
6. Temperature and humidity calibration
SK-110TRH CALIBRATION
1

Advertisement

loading

Summary of Contents for SATO SK-110TRH

  • Page 1 SK-110TRH CALIBRATION To calibrate the thermohygrometer, use a temperature/humidity generating chamber to produce reference temperature and reference humidity. A chamber capable of achieving a temperature repeatability within ±0.1°C, humidity repeatability within ±1.0%, and humidity variability between 0 and 90% is required.
  • Page 2 Temperature calibration The upper figure of the LCD indicates temperature and the lower figure indicates humidity. The value is decreased by turning the respective volume control counterclockwise and increased by turning it clockwise. (1) Leave at 23°C, 0%, for 30 minutes. (2) Use a flat head screwdriver to adjust VR1 to 23.0°C.
  • Page 3 7. Replacing the front case The screws that fasten the two sections of the case together are for plastic. Tighten the screws in the order shown to a torque of 5 kgf/cm...