Documentation; Test Equipment - National Instruments NI-9269 Calibration Procedure

Hide thumbs Also See for NI-9269:
Table of Contents

Advertisement

Documentation

Consult the following documents for information about the NI 9269, NI-DAQmx, and your
application software. All documents are available on
software.
NI cDAQ-9174/9178 Quick Start
NI-DAQmx installation and hardware setup
NI 9269 Operating Instructions and Specifications
NI 9269 specific information, specifications, and calibration interval
NI-DAQmx Readme
Operating system and application software support in NI-DAQmx
LabVIEW Help
LabVIEW programming concepts and reference information about NI-DAQmx
VIs and functions
NI-DAQmx C Reference Help
Reference information for NI-DAQmx C functions and NI-DAQmx C properties
NI-DAQmx .NET Help Support for Visual Studio
Reference information for NI-DAQmx .NET methods and NI-DAQmx .NET
properties, key concepts, and a C enum to .NET enum mapping table

Test Equipment

Table 1 lists the equipment recommended for the performance verification and adjustment
procedures. If the recommended equipment is not available, select a substitute using the
requirements listed in Table 1.
Equipment
DMM
Chassis
2 | ni.com | NI 9269 Calibration Procedure
Table 1. Recommended Equipment
Recommended Model
NI 4070 DMM
NI cDAQ-9178
and help files install with the
ni.com
Requirements
Use a multiranging 6 1/2 digit
DMM with a voltage accuracy of
±40 ppm.