Hanna Instruments HI98198/10 Instruction Manual page 54

Optical dissolved oxygen meter
Table of Contents

Advertisement

54
Find Quality Products Online at:
1. Unplug the
HI764113
from the
HI98198
steel guard from probe body and set aside. Wipe off the probe and cap
with a soft tissue.
2. Remove the expired "Smart Cap™" from the probe; Squeeze the used
cap at the notched V and pull it off the probe body (without twisting).
3. Remove the used O‑ring from the O‑ring groove by rolling it off the probe.
4. Wipe the O‑ring groove with a soft tissue followed by a cleaning with
the supplied lens cleaner wipe.
5. Remove the new O‑ring from the
body using care not to roll or twist the O‑ring.
6. Sparingly lubricate the O‑ring with a thin film of the supplied grease. Use
care to prevent grease or fingerprints from contacting the optical window.
7. Clean the lens with an unused portion of the supplied lens cleaner wipe.
8. Remove the new optical cap from the box. Align the notched cutout
arrow on the Smart Cap™ with the matching guide on the probe body.
9. Slide and press the Smart Cap™ onto the
cap snaps in place. Once the cap is installed, it should not be removed
unless a new cap is required.
10. Connect the
HI764113
probe to the
the connector to the DIN jack located on the top of the meter.
11. Power the meter to initiate the cap timer.
12. Calibrate.
13. Reinstall the stainless steel guard onto the probe body.