insensitive material. Please do not use it if any person is allergic to lanyard. Do not
wrap the lanyard around neck to avoid an accident.
M The disposal of scrap device, its accessories and packaging should follow the local
laws and regulations, to avoid polluting to the local environment. And the packaging
materials must be placed in the region where the children are out of reaching.
M The device can not be used with the equipment not specified in the Manual. Only the
accessories appointed or recommended by the manufacturer can be used, otherwise
it may cause injury to the tester and operator or damage to the device.
M Check the device before use to make sure that there is no visible damage that may
affect user's safety and device performance. When there is obvious damage, please
replace the damaged parts before use.
M Functional testers can not be used to assess the accuracy of the Pulse Oximeter.
M Some functional testers or patient simulators can be used to verify whether the device
works normally, for example, INDEX-2LFE Simulator (software version: 3.00), please
refer to the Manual for the detailed operation steps.
M Some functional testers or patient simulators can measure the accuracy of the device
copied calibration curve, but they can not be used to evaluate the device accuracy.
M When using the device, please keep it away from the equipment which can generate
strong electric field or strong magnetic field. Using the device in an inappropriate
environment may cause interference to the surrounding radio equipment or affect its
working.
M When storing the device, keep it away from children, pets and insects to avoid affecting
its performance.