Recertification; How Often To Recertify; Where To Send A Kit For Recertification - Agilent Technologies 85057B User's And Service Manual

2.4 mm verification kit
Table of Contents

Advertisement

Recertification

The following will be provided with a recertified kit:
• a new calibration sticker affixed to the case
• a certificate of calibration
• a calibration report for each device in the kit listing measured values, specifications,
and uncertainties
A list of NIST traceable numbers may be purchased upon request to be
NOTE
included in the calibration report.
Agilent Technologies offers a Standard calibration for the recertification of the kit. For
more information, contact Agilent Technologies. Refer to

How Often to Recertify

The suggested initial interval for recertification is 12 months or sooner. The actual need for
recertification depends on the use of the kit. After reviewing the results of the initial
recertification, you may establish a different recertification interval that reflects the usage
and wear of the kit.
NOTE
The recertification interval should begin on the date the kit is first used after
the recertification date.

Where to Send a Kit for Recertification

Contact Agilent Technologies for information on where to send your kit for recertification.
Refer to
"Contacting Agilent" on page
Technologies" on page 5-4
85057B
5-5. Refer to
for details on sending your kit.
"Contacting Agilent" on page
"Returning a Kit or Device to Agilent
Performance Verification
Recertification
5-5.