7. Disconnect the short and gage from the airline:
If the airline center conductor does not disengage from the device center
NOTE
conductor, gently pull the center conductors apart and then push the airline
center conductor back inside the outer conductor of the airline.
a. Without turning the airline or the gage, remove the gage from the airline. Refer to
"How to Separate a Connection" on page
b. Replace the protective end cap on the airline to prevent the center conductor from
sliding out of the outer conductor.
c. Without turning the airline or the short, remove the short from the airline. Refer to
"How to Separate a Connection" on page
d. If you will not be using the airline again immediately, slide the center conductor out
of the outer conductor and store the center conductor in the plastic case provided.
e. Replace the other protective end cap on the outer conductor and store the center and
outer conductors in the foam lined storage case.
85057B
Use, Maintenance, and Care of the Devices
3-18.
3-18.
Gaging Connectors
3-13