Invacare Platinum Mobile User Manual page 74

Oxygen concentrator
Hide thumbs Also See for Platinum Mobile:
Table of Contents

Advertisement

Invacare® Platinum® Mobile
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Device is used exceeds the applicable RF compliance level above, the Device
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the Device.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
At 80 MHz and 800 MHz, the higher frequency range applies.
Recommended separation distances between portable and mobile RF communications equipment and the Device
The Device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Device can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Device as recommended below, according to the
maximum output power of the communications equipment
Rated maximum output
of transmitter [W]
0.01
0.1
1
10
100
74
Separation distance according to frequency of transmitter [m]
150 kHz to 80 MHz
d = 1.2√P
0.12
0.37
1.17
3.69
11.67
80 MHz to 800 MHz
d = 0.35√P
0.04
0.11
0.35
1.11
3.50
800 MHz to 2,5 GHz
d = 0.7√P
0.07
0.22
0.70
2.21
7.00
1194969-C