Technical Information
Table 25
Offset
Meaning
7
Interface Status
B
→ Controller not responds
0
B
→ Not used
1
B
→ Input data not fully coherent yet
2
B
→ Not used
3
B
→ Not used
1...7
8
B
→ HV1 On/Off diagnostic status
7...4
B
→ HV2 On/Off diagnostic status
3...0
9
B
→ HV3 On/Off diagnostic status
7...4
B
→ HV4 On/Off diagnostic status
3...0
10
B
→ HV1 Protect Mode diagnostic status
7...4
B
→ HV2 Protect Mode diagnostic status
3...0
11
B
→ HV3 Protect Mode diagnostic status
7...4
B
→ HV4 Protect Mode diagnostic status
3...0
12
B
→ HV1 Fixed/Step Mode diagnostic status
7...4
B
→ HV2 Fixed/Step Mode diagnostic status
3...0
13
B
→ HV3 Fixed/Step Mode diagnostic status
7...4
B
→ HV4 Fixed/Step Mode diagnostic status
3...0
14
B
→ Serial/Remote/Local/Profibus diagnostic status
7...4
B
→ Not used
3...0