Agilent Technologies 4UHV User Manual page 210

Ion pump controller
Hide thumbs Also See for 4UHV:
Table of Contents

Advertisement

Technical Information
Command Listing
It is not possible to implement all the "Binary Dual" protocol commands on the
4UHV controller. The commands that are implemented on the 4UHV controller are
the following:
Table 16
No.
1
3
4
5
7
8
9
10
11
12
13
14
15
16
17
18
19
210
Description
Local/remote
HV-ON/OFF
Device number
Measurement unit
Protect
Fixed/Step
Power supply polarity
Max. power
Vmax
Iprotect
Set Point
V meas.
I meas.
P meas.
Set Point status
Interlock status
Error status
ASCII
Z0
A0
F0
D0
C0
B0
G0
J0
H0
K0
P0
S0
T0
U0
g0
]0
z0
Agilent 4UHV Ion Pump Controller User Manual
Hex
5Ah 30h
41h 30h
46h 30h
44h 30h
43h 30h
42h 30h
47h 30h
4Ah 30h
48h 30h
4Bh 30h
50h 30h
53h 30h
54h 30h
55h 30h
67h 30h
5Dh 30h
7Ah 30h

Advertisement