The determined values, including the measuring points, must be recorded and assessed
in the protective conductor resistance report.
The measuring procedure and the measuring device used (designation and serial num-
ber) are also to be documented.
1.5.7

Information on measuring the system leakage current

NOTE
Siemens AG
Medical Solutions
Measuring circuit for measuring the protective conductor resistance in units/systems that
are permanently connected to the power supply net (according to DIN VDE
0751-1:2001-10, Fig. C3).
System
Application part (not available)
Measurement setup (integrated into measuring device)
The system leakage current measurement is to be conducted and
recorded as the repeat measurement during maintenance.
However, the first measured value must be newly determined and
a new report must created under the following conditions:
Lack of system leakage current measurement documentation
Local line voltage or line frequency deviating from the line voltage
and line frequency documented in the report (e.g. in the event of a
site/operator change)
Use of a different procedure for measuring the system leakage cur-
rent from the one documented in the report.
For the purpose of traceability, reference to the new report is to be
written on the old report. The reason for newly determining the
first measured value is to be documented and confirmed with a
name and signature.