6.3

Performance Properties

Fluorescence LED illumination system
LED main wavelength
Optical multiplexing
method
Mounting section for the
epi fluorescence
attachment
Pillar attaching part
Automatic wavelength
recognition
Cooling method
Connection
Field number
On/off and
brightness control
From the remote
control pad
From NIS-Elements Selection of LEDs and turning them on/off:
From the Ti2-E Can be controlled by using the operation buttons on the microscope main body, the
Chapter 6 Specifications
Four types (385 nm, 475 nm, 550 nm, and 621 nm)
Supports simultaneous/individual illumination control and individual adjustment
Built-in dichroic mirror
Dedicated mount:
Two M5 x 8 mm holes, on the top and bottom of the housing
Measures wavelength information from LED units, and displays the LED center
wavelength (Typ.) in NIS-Elements.
Air cooled heat sink inside main body
Install this product with its top or bottom vent facing upwards.
USB connector:
REMOTE connector:
CONTROL connector:
DC connector:
TRIGGER connector:
22 (25 when the TI2-LA-FL-2 FL module is used)
Switches between all-LED simultaneous control (SYNC mode) and individual
control (ASYNC mode)
Selection of LEDs and turning them on/off:
Illumination control step:
Illumination control step:
tablet application Ti2 Control, and the TI2-S-JS joystick controller.
Selection of LEDs and turning them on/off:
Illumination control step:
Controlling the power:
Other:
• In the all-LED simultaneous control mode, when multiple LEDs are selected, the
brightness values of each wavelength can be controlled while maintaining
relative brightness between wavelengths.
• Excitation light emission synchronized to the image capture timing on the CCD
camera (time-lapse observation)
• Supports the trigger acquisition function
30
28 mm-diameter, 13 mm-deep, with an anti-rotation
block
Fix with set screws (2 mm across flats) from the holes
on the left and right sides of the main body.
For connecting a PC, USB 2.0 B connector
For connecting the remote control pad (14-pin)
For connecting the Ti2-E (14-pin)
For connecting the AC adapter
For trigger input/status output (high density D-Sub 15
pin, female)
Multiple LEDs can be turned on simultaneously;
simultaneous control is available for single color only.
0 to 100% (101 steps for each LED)
Multiple LEDs can be turned on simultaneously;
simultaneous control is available for single unit only.
0 to 100% (101 steps for each LED)
Multiple LEDs can be turned on simultaneously;
simultaneous control is available for single unit only.
0 to 100% (101 steps for each LED)
Interlocks with the power condition of the Ti2
microscope main body.
The FL turret address and the filter cube on the optical
path can be displayed.

Hide quick links: