Activates the span functions menu. The menu selections are determined by the current
measurement mode.
Mode
Preset
State Saved
Key Path
Span
Sets the span frequency value. Span frequency describes the frequency range that is
positioned symmetrically about center frequency.
Changing span changes the displayed frequency range symmetrically about the center
frequency.
While in the SA measurement, setting the span to 0 Hz through SCPI or the front-panel
numeric key pad puts the analyzer into the Zero Span Measurement.
While in the Zero Span measurement, setting the Span to a non-zero value through SCPI
or Front Panel puts the analyzer in the SA measurement.
In a Zero Span measurement, the frequency entry mode is always center/span.
Mode
Remote Command
Example
Dependencies /
Couplings
Remote Command
Notes
Preset
State Saved
76
All
Full span (depends on hardware present – options)
Saved in State
Front-panel key
SA
[:SENSe]:FREQuency:SPAN <frequency>
[:SENSe]:FREQuency:SPAN?
FREQ:SPAN 2 GHz sets the span to 2 GHz
FREQ:SPAN 0 Hz sets the span to 0 Hz and puts the
instrument in Zero Span measurement mode.
Span affects RBW
Preset and Max values are dependent on Hardware (Options)
2.9999 GHz (Option 503)
5.9999 GHz (Option 506)
Saved in State
Chapter 3