Special instructions:

STORAGE CONDITIONS:

Recommended shipping and storage at 5°C - 40°C, non-condensing
Avoid direct sunlight
Maintenance and Cleaning: The Monitor has a useful life of 5 years. The Monitor contains no user
serviceable components and requires no maintenance or calibration. Keep it clean and free of dust.
The exterior may be cleaned using the following steps:
1. After every use, check the Monitor for any sign of damage or wear. Remove any external lead still
connected in the A or B ports and discard per local procedures.
2. Wipe the Monitor with a dry or water-moistened soft cloth, Isopropyl Alcohol, Ammonium
Hydroxide based surface cleanser, Ammonium Chloride based surface cleanser or 2% bleach solution.
Ensure any residual organic material is removed. Do not pour or spray liquid directly on the Monitor.
Allow to air dry before use.
3. Do not put liquid near the speaker.
4. The FLOW COUPLER Monitor is no longer usable when hardware buttons, LCD screens are
inoperable, or audible signal is not generated when connected to the FLOW COUPLER probe.
The Following Sections apply to the FLOW COUPLER 20 MHz Doppler Probe
Acoustic Output: There are no user controls meant to a ect acoustic outputs. All acoustic outputs are
below the application speci c pre-amendments acoustical output limit of an Ispta of 94 mW/cm
a MI of 1.9.
Explanation of Derivation of Derating Factor: The derated intensity calculations are based on
measured center frequency of the acoustical signal (f, MHz) and the distance from the transducer
under test to the hydrophone (z, cm) using the derating factor e.
Track 1 summary
System: FLOW COUPLER Monitor and Probe System
Monitor: FLOW COUPLER Monitor
Mode of Operation
Clinical Application
B
M
Ophthalmic
Other (intra-operative
and post-operative)
Cardiac
Clinical Application
8.
-0.069 fz
PWD
CWD
Color
Combined
Doppler
(Specify)
X
Symbol Index
PWD:
MI
TIS
TIB
TIC
P
r,a
P
P
1x1
z
s
z
b
Z
MI
2
and
Z
pii,a
f
awf
prr
srr
n
pps
I
pa,a
I
Other
spta,α
(Specify)
z
sii,α
I
spta
z
pii
z
sii
p
r
Pulsed Wave Doppler
Mechanical Index
Soft Tissue Thermal Index
Bone Thermal Index
Cranial Bone Thermal Index
Attenuated Peak-Rarefractional Acoustic Pressure
Power Output
Power Output through 1cm
2
of area
Depth for TIS
Depth for TIB
Depth for Mechanical Index
Depth for Peak Attenuated Pulse Intensity Integral
Acoustic Working Frequency
Pulse Repetition Rate
Scan Repetition Rate
Number of Pulses per Ultrasonic Scan Line
Attenuated Pulse-Average Intensity
Attenuated Spatial-Peak Temporal Average Intensity
Depth for Peak Sum of Attenuated Pulse
Spatial-Peak, Temporal-Average Intensity
Depth for Peak Pulse-Intensity Integral
Depth for Peak Sum of Pulse-Intensity Integral
Peak-Rarefractional Acoustic Pressure