Figure 4-9: mA/HART multidrop wiring
A. Input Device 1
B. Input Device 2
C. Input Device 3
D. Meter (mA+/HART output)
E. Hazardous area
F. Non-hazardous area
G. DIN rail
H. 250 Ω resistance
I. 24VDC
J. HART/Field Communicator
4.11
Wire the Frequency Output or the Discrete
Output
Use this procedure to wire Channel C or D as the Frequency Output or the Discrete Output.
Wire the FO or DO in explosion-proof, intrinsically safe, or nonhazardous installations.
Meter installation and wiring should be performed only by suitably-trained personnel
using the appropriate government and corporate safety standards.
Procedure
Wire to the appropriate output terminal and pins.
Installation Manual
WARNING
Wiring the channels
October 2022