Use and Operation

Important advise for storage

• The goods should only be stored in the provided
drawers and/or racks.
• The loading must never obstruct the air-inlet-
openings inside the unit.

Alarm situations

• On the occurrence of an alarm situation attempts
must be made to find out the reason for the
alarm and remove it as quickly as possible. If that
is not successful the necessary measures must
immediately be taken so that the stored goods do
not sustain any damage.
• At a temperature alarm the red LED alarm lights
up and an acoustic signal sounds. In addition the
corresponding alarm message appears on the
display.
• In the case of a powerfail alarm the green power
LED and the red LED alarm flash and an acoustic
signal sounds. The display is switched off but be
temporarly switched on again by pressing any
key.

Alarm History

• The alarm history list contains all relevant data
on temperature and powerfail events. The list
contains up to 20 alarm situations.
• The status message "HISTORY" indicates that
new entries have been made in the list.
• The following information is stored in the Alarm History :
w
Temperature alarm :
ALARM :
START :
END :
AVG :
MAX / MIN :
Configuration fault: Fault in the parameter setting. The appliance does
not start.
Electronics fault. The function of the appliance is stopped.
No communication between interface and mainboard. The function of
the appliance is stopped.
Serious fault, however the appliance continues to run in "FAILSAFE"
mode. i.e. the compressor switches on and off according to the set
rhythm.
type of alarm (HIGH ALARM or LOW ALARM)
Start date and time of alarm situation
End date and time of alarm situation
Average temperature on the display sensor during the alarm situations.
Maximum or minimum temperature value on the display sensor during the
alarm situation (depending on the type of temperature alarm)
• Make sure that the loaded goods do not tilt to the
backside.
• Make sure that the loading does not touch the
reference bottle of the display sensor.
• Do not overload the unit.
After a given time (setting: EXTRA - DISPLAY -
BACKL. TIMEOUT ) the display switches off again.
The acoustic alarm can be suppressed for a set
time by pressing the ENTER button. During a
powerfail phase no parameters can be changed.
• If the internal temperature rises above the upper
limit during a powerfail phase a temperature
alarm is triggered and entered into the History
list. In the case of a door alarm the red LED lights
up and an acoustic signal sounds. In addition the
corresponding alarm message appears on the
display
• The alarm history list can be accessed via the
"SHOW HISTORY" function in the EXTRA menu.
Within the list you navigate by means of the UP
and DOWN buttons. The first and last entry in the
list are marked "BEGIN OF LIST" and "END OF
LIST"