National Instruments USB-621 Series User Manual page 61

Hide thumbs Also See for USB-621 Series:
Table of Contents

Advertisement

Chapter 4
Analog Input
AI Ground-Reference
*
Setting
Differential (DIFF)
Non-Referenced
Single-Ended (NRSE)
Referenced Single-Ended
(RSE)
*
Refer to the
Analog Input Ground-Reference Settings
software considerations.
Refer to the
Connecting Ground-Referenced Signal Sources
Note: (USB-621x BNC Devices) Refer to the appropriate Analog Input section for your USB-621x BNC device in
Appendix A,
Device-Specific
device.
NI USB-621x User Manual
Table 4-2. USB-6210/6211/6212 Analog Input Configuration
Floating Signal Sources
(Not Connected to
Building Ground)
Examples:
• Ungrounded thermocouples
• Signal conditioning with
isolated outputs
• Battery devices
Signal Source
+
Signal Source
+
Signal Source
+
Information, for more information about using grounded and floating signal sources with your
Example:
• Plug-in instruments with
DAQ Device
AI+
+
AI–
AI GND
DAQ Device
AI
+
AI SENSE
AI GND
DAQ Device
AI
+
AI GND
section for descriptions of the RSE, NRSE, and DIFF modes and
section for more information.
4-24
Ground-Referenced
Signal Sources
non-isolated outputs
DAQ Device
Signal Source
AI+
+
AI–
AI GND
V cm
Signal Source
DAQ Device
AI
+
AI SENSE
AI GND
V cm
NOT RECOMMENDED for the
USB-6210/6211/6212
Signal Source
DAQ Device
AI
+
V
A
AI GND
V
B
Ground-loop potential (V
– V
) are added
A
B
to measured signal.
+
+
+
ni.com

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Usb-6210Usb-6211Usb-6212Usb-6215Usb-6216Usb-6218

Table of Contents