Assembly - Siemens Mobilett Plus Installation Instructions Manual

Hide thumbs Also See for Mobilett Plus:
Table of Contents

Advertisement

Installation

Assembly

Fig. 1
Removal of Mobilett covers
1.
2.
3.
NOTICE
Siemens-Elema AB
Solna, Sweden
Remove the Mobilett top cover plate (pos 1 / Fig. 1), and the protection plate of the
CPU board below, to expose the main CPU D1, see Fig. 6.
Remove the cover of the Mobilett articulated arm (pos 2 / Fig. 1), the right and left
side covers (pos 3 / Fig. 1), both the right and left circular covers (pos 4 / Fig. 1),
the rear cover of the articulated shaft (pos 5 / Fig. 1) and the front cover with
cassette holder (pos 7 / Fig. 1).
Switch S10 to discharge position. After approximately 5 minutes the capacitors
will be discharged. Measure the discharge procedure between measuring points
CAPPOS and CAPNEG on the D7 board (pos 6 / Fig. 1) with a digital multimeter.
For MOBILETT Plus HP only:
Remove the cover in front of the batteries. Disconnect
one of the blue battery connector pairs.
2
5
8
6
SPR8-220.031.02
Rev. 02
11.02
2
4
3
Page 1 of 8
Mobilett Plus/ E/ M/ HP
SPS-UD
2 - 1
2
1
M
C
B
7
Service