Siemens MOBILETT XP Troubleshooting Manual page 38

Hide thumbs Also See for MOBILETT XP:
Table of Contents

Advertisement

38
ERR 9
System
mode:
Standby
ERR 10
System
mode:
Standby
ERR 11
System
mode:
Standby
MOBILETT XP
Static kV measured value (tube current) <> 0
Explanation: A tube current is detected during stand-by. D916 (CPU)
detects actual mA error.
Action:
Check the cable connection between measurement data log D907 (the
single tank) and D916.X8 (CPU).
Possible causes: Cable connection between the single tank and D916 (X8);
"ground connection" from the system to the single tank is defective; D916;
D907; single tank.
Defect in a battery pack
Explanation: D916 (CPU) detects via D982 a battery pack with a voltage <=
125V or a defective battery pack.
Action:
Check the battery voltage, nominal value > 125 V, possibly charge bat-
tery.
Check fuse F5 on D982 (battery block output).
Check fuse F3 on D982 (charger 1).
Check fuse F4 on D982 (charger 2).
Check the cable connection between the battery blocks and D982.
Check the cable/contacts between the individual batteries.
Capacitor bank voltage too high
Explanation: D916 (CPU) detects voltage too high (> 460 V) on the capaci-
tor bank.
Action:
Check measurement point TP-VC on D916 (capacitor bank voltage)
Possible causes: D952
Rare causes: Cable connection X20 (D916-D952-D962); D916
SPR8-230.840.01.05.02
11.05
Error messages
Page 38 of 72
CS SD 24
Siemens AG
Medical Solutions