MOBILETT XP
Static kV measured value (tube current) <> 0
Explanation: A tube current is detected during stand-by. D916 (CPU)
detects actual mA error.
Action:
•
Check the cable connection between measurement data log D907 (the
single tank) and D916.X8 (CPU).
Possible causes: Cable connection between the single tank and D916 (X8);
"ground connection" from the system to the single tank is defective; D916;
D907; single tank.
Defect in a battery pack
Explanation: D916 (CPU) detects via D982 a battery pack with a voltage <=
125V or a defective battery pack.
Action:
•
Check the battery voltage, nominal value > 125 V, possibly charge bat-
tery.
•
Check fuse F5 on D982 (battery block output).
•
Check fuse F3 on D982 (charger 1).
•
Check fuse F4 on D982 (charger 2).
•
Check the cable connection between the battery blocks and D982.
•
Check the cable/contacts between the individual batteries.
Capacitor bank voltage too high
Explanation: D916 (CPU) detects voltage too high (> 460 V) on the capaci-
tor bank.
Action:
•
Check measurement point TP-VC on D916 (capacitor bank voltage)
Possible causes: D952
Rare causes: Cable connection X20 (D916-D952-D962); D916
SPR8-230.840.01.05.02
11.05
Error messages