Mount The Process Seal Antenna - Emerson Rosemount 3408 Reference Manual

Level transmitter, non-contacting radar
Hide thumbs Also See for Rosemount 3408:
Table of Contents

Advertisement

Reference Manual
00809-0100-4418
3.4

Mount the process seal antenna

3.4.1
Mount the flanged version
Procedure
1. Lower the transmitter into the nozzle.
2. Tighten the bolts and nuts (see
3.4.2
Torque specifications
The conditions used for the calculation are:
Rosemount 3408 Level Transmitter
Note
Be careful not to scratch or otherwise damage the PTFE sealing.
Note
Re-tighten after 24 hours and again after the first temperature cycle.
Check at regular intervals and re-tighten if necessary.
Table
3-5).