13 - 4

Filament circuit faults

When a fault occurs in the filament circuit, the following actions are initiated:
The buzz on board D3
generates 10 beep tones,
the green "ready" LED on board D3 is turned off,
the Control and display panel is switched off,
the filament current is switched off,
an error message is shown in the Control and display panel.
The error message can be acknowledged by pressing the "kV+" selection button.
If this fails to switch off the error message, proceed according to the Service Instructions!
MOBILETT Plus/ Plus E
Register 6
SPR8-215.850.02
Rev. 01
Page 4 of 4
01.02
SPS-UD
Filament circuit
Siemens-Elema AB
Solna, Sweden