Sonication Time - BANDELIN SONOREX SUPER RK 31 Operating Manual

Hide thumbs Also See for SONOREX SUPER RK 31:
Table of Contents

Advertisement

5/
Operation
• Telephone advice: +49 30 76880-280
• Internet: www.dr-stamm.de
Observe the information on dosing provided by the ultrasound
agent manufacturer. Please refer to the dosing table for the
required quantities of ultrasound agent and water. See chapter
9 Dosing table.
You can calculate the quantities yourself in the same way as the
following example:
10 l ready to use solution, 2.5%:
10 l x 2,5 %
100 %
10 l - 0,25 l = 9,75 l water
5.3

Sonication time

NOTICE
Risk of damage to the goods to be treated
Excessively long sonication can damage the surface of the goods to
be treated.
– Select a short sonication time where possible.
The ideal sonication time depends on several factors:
• Type and concentration of agent
• Working temperature of sonication fluid
• Type of soiling
• Type of goods to be treated, in particular materials
Note the information on the recommended sonication time from
the agent manufacturer. To begin with, select as short a sonication
time as possible to protect the objects to be sonicated and the
ultrasonic oscillating tank. Check the result. Lengthen the sonica-
tion time if the result is not sufficient.
20/52
= 0,25 l agent
1676-002 en/2022-05